Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Carestream Health, Inc. recalls The DRX-Revolution Mobile X-Ray System is a diagnostic mobile x-ray system utilizing digital radiography (DR)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
There is a potential for unexpected failure of the electrical components within the Carestream Health Inc. generator installed in your DRX-Revolution Mobile X-ray System(s). This can lead to temporary and self contained thermal overload within the generator. Should an event like this occur, the system will become inoperable and loud noise, burnt smell and smoke may be detectable.
Products
  • The DRX-Revolution Mobile X-Ray System is a diagnostic mobile x-ray system utilizing digital radiography (DR)
Where it was sold

US Nationwide. Global Distribution.

Nationwide

Barcodes
6088997101939460889978618897
Units affected
2402 systems
About this recall
The DRX-Revolution Mobile X-Ray System is a diagnostic mobile x-ray system utilizing digital radiography (DR) technology. the system consists of a self-contained x-ray generator, image receptor(s), imaging display and software for acquiring medical diagnostic images outside of standard stationary x-ray room. The DRX-Revolution system incorporates a flat-panel detector that can be used wirelessly for exams such as in-bed chest projections. The system can also be used to expose CR phosphor screens or films.