Philips Respironics, Inc. recalls Trilogy Evo. Model Numbers: DS2110X11B
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
- Products
-
- Trilogy Evo. Model Numbers: DS2110X11B
- Company
- Philips Respironics, Inc.
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, LY, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, TZ, UA, UY, VN, YT, ZA, ZW.
Nationwide
- Units affected
- 82,800 units
- About this recall
- Trilogy Evo. Model Numbers: DS2110X11B, UDS2110X11B, RDS2110X11B, CA2110X12B, JP2110X16B, IN2110X15B, DE2110X13B, AU2110X15B, BL2110X15B, EU2110X15B, BR2110X18B, LA2110X15B, EE2110X15B, FR2110X14B, GB2110X15B, ND2110X15B, ES2110X15B, IA2110X15B, IT2110X21B, KR2110X15B, TR2110X15B, EU2110X19, FP2110X10, PP2110X10, RBRBR2110X18B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.