Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Ge Medical Systems Israel, Functional Imaging recalls Elscint Model 4XX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Products
  • Elscint Model 4XX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Where it was sold

Worldwide

Nationwide

Units affected
64 units
About this recall
Elscint Model 4XX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging