Philips North America Llc recalls Big Bore RT with software version V4.8.0.10421- Model: 728242 CT scanner and simulator designed for radiation
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.
- Products
-
- Big Bore RT with software version V4.8.0.10421- Model: 728242 CT scanner and simulator designed for radiation
- Company
- Philips North America Llc
- Where it was sold
-
Nationwide Foreign: Austria Belgium Canada China Estonia France Germany Hondorus India Italy Mexico Netherlands Philippines Poland Singapore Spain Switzerland Taiwan
Nationwide
- Barcodes
00884838095168- Units affected
- 56 units
- About this recall
- Big Bore RT with software version V4.8.0.10421- Model: 728242 CT scanner and simulator designed for radiation oncology and therapy.