Philips North America Llc recalls Brilliance CT Big Bore Oncology with software version V4.8.0.10421 Model:728243 CT scanner and simulator desig

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.
Products
  • Brilliance CT Big Bore Oncology with software version V4.8.0.10421 Model:728243 CT scanner and simulator desig
Where it was sold

Nationwide Foreign: Austria Belgium Canada China Estonia France Germany Hondorus India Italy Mexico Netherlands Philippines Poland Singapore Spain Switzerland Taiwan

Nationwide

Barcodes
00884838059450
Units affected
29 units
About this recall
Brilliance CT Big Bore Oncology with software version V4.8.0.10421 Model:728243 CT scanner and simulator designed for radiation oncology and therapy.