Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Microbiologics Inc recalls KWIKSTIK

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
A complaint investigation showed that the bulk pellets (lot 794-142) were contaminated with Aspergillus flavus.
Products
  • KWIKSTIK
Where it was sold

Worldwide distribution - US Nationwide and the countries of Italy, France, Turkey, Peru, Mexico.

Nationwide

Barcodes
2084535701916930845357019173
Units affected
13 units
About this recall