Boston Scientific Neuromodulation Corporation recalls Nalu TD2 Therapy Disc with firmware
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Therapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a firmware issue that causes the TD to attempt to restart, then fail, leading it to become non-functional, and if a secondary TD is unavailable, this may lead IPG to become non-functional, thus unable to inhibit transmission of pain signals.
- Products
-
- Nalu TD2 Therapy Disc with firmware
- Where it was sold
-
US Nationwide distribution in the states of PA, FL, OR, AZ, OH, SC, TX, OR, OK, MI, CA, NJ.
States: Arizona, California, Florida, Michigan, New Jersey, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, Texas
- Units affected
- 134
- About this recall
- Nalu TD2 Therapy Disc with firmware, REF: 34014-001, and TD2 Patient Kit, REF: 74011, part of the Nalu Neurostimulation System