Beckman Coulter, Inc. recalls The Access Cortisol assay is a paramagnetic particle
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The Access Cortisol assay protocol file (APF) applies an incorrect calibration curve acceptance parameter at the S1 calibrator level that can lead to a high rate of calibration failures which can vary by reagent lot number. Calibration failure would lead to delay of measurement results reporting and potential injuries.
- Products
-
- The Access Cortisol assay is a paramagnetic particle
- Company
- Beckman Coulter, Inc.
- Where it was sold
-
Worldwide distribution.
Nationwide
- Barcodes
15099590231118- Units affected
- 149 units
- About this recall
- The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.