Beckman Coulter, Inc. recalls The Access Cortisol assay is a paramagnetic particle

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The Access Cortisol assay protocol file (APF) applies an incorrect calibration curve acceptance parameter at the S1 calibrator level that can lead to a high rate of calibration failures which can vary by reagent lot number. Calibration failure would lead to delay of measurement results reporting and potential injuries.
Products
  • The Access Cortisol assay is a paramagnetic particle
Where it was sold

Worldwide distribution.

Nationwide

Barcodes
15099590231118
Units affected
149 units
About this recall
The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.