Surgical Theater Inc recalls SuRgical Planner (SRP)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.
- Products
-
- SuRgical Planner (SRP)
- Company
- Surgical Theater Inc
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.
States: Arizona, California, District of Columbia, Florida, Indiana, Kentucky, Michigan, Minnesota, New Jersey, New York, Pennsylvania, Texas, West Virginia
- Barcodes
00863157000336- Units affected
- 45 units
- About this recall
- SuRgical Planner (SRP), software version SRP.7.12.0, imaging processing radiological system