Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Natus Medical Incorporated recalls Embletta MPR PG Multi Parameter Recorder Sleep Monitor

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The device may be unable to correctly format its internal memory. When the internal memory cannot be correctly formatted, previously recorded studies may remain on the device and may be updated with incorrect patient information from a subsequent study. The issue may also prevent studies from initiating recording successfully or may cause recordings to end prematurely, resulting in incomplete or unavailable diagnostic data and the potential need to repeat a patient study.
Products
  • Embletta MPR PG Multi Parameter Recorder Sleep Monitor
Where it was sold

Worldwide distribution - US Nationwide and the countries of Canada, China, Italy, Portugal.

Nationwide

Barcodes
00382830023764
Units affected
288 units
About this recall
Embletta MPR PG Multi Parameter Recorder Sleep Monitor, Model/Catalog Number: 2003011