Medtronic Inc recalls Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect identification of the proximal and distal inflation /deflation ports when viewing the devices from the inferior side.
- Products
-
- Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device
- Company
- Medtronic Inc
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Algeria, Armenia, France, Germany, Iraq, Italy, Japan, Netherlands, Poland, Romania, Spain, Sweden, Switzerland, United Kingdom.
Nationwide
- Barcodes
0803347709071900763000451691- Units affected
- 2767 units
- About this recall
- Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MUS0130069X6