Baxter Healthcare Corporation recalls Hillrom Welch Allyn HScribe Holter System; ambulatory monitoring

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.
Products
  • Hillrom Welch Allyn HScribe Holter System; ambulatory monitoring
Where it was sold

Worldwide distribution.

Nationwide

Barcodes
007320942639920073209426404300732094264074007320942641110073209429629700817655020112008176550201290081765502022800817655020365008176550204260081765502174400817655022161
Units affected
599 units
About this recall
Hillrom Welch Allyn HScribe Holter System; ambulatory monitoring