Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Baxter Healthcare Corporation recalls Hillrom Welch Allyn RScribe 12-Lead Electrocardiograph System; Electrocardiograph

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.
Products
  • Hillrom Welch Allyn RScribe 12-Lead Electrocardiograph System; Electrocardiograph
Where it was sold

Worldwide distribution.

Nationwide

Barcodes
0073209426295700732094263046007320942631070073209426313800732094263152007320942631690081234502782200812345027839008176550202800081765502136200817655022246
Units affected
399 units
About this recall
Hillrom Welch Allyn RScribe 12-Lead Electrocardiograph System; Electrocardiograph