Baxter Healthcare Corporation recalls Hillrom Welch Allyn RScribe 12-Lead Electrocardiograph System; Electrocardiograph
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.
- Products
-
- Hillrom Welch Allyn RScribe 12-Lead Electrocardiograph System; Electrocardiograph
- Where it was sold
-
Worldwide distribution.
Nationwide
- Barcodes
0073209426295700732094263046007320942631070073209426313800732094263152007320942631690081234502782200812345027839008176550202800081765502136200817655022246- Units affected
- 399 units
- About this recall
- Hillrom Welch Allyn RScribe 12-Lead Electrocardiograph System; Electrocardiograph