Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Baxter Healthcare Corporation recalls Hillrom Welch Allyn Diagnostic Cardiology Suite (formerly Connex Cardio); Diagnostic spirometer

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.
Products
  • Hillrom Welch Allyn Diagnostic Cardiology Suite (formerly Connex Cardio); Diagnostic spirometer
Where it was sold

Worldwide distribution.

Nationwide

Barcodes
0081765502259800817655022611
Units affected
1369 units
About this recall
Hillrom Welch Allyn Diagnostic Cardiology Suite (formerly Connex Cardio); Diagnostic spirometer