Baxter Healthcare Corporation recalls Hillrom Welch Allyn XScribe Cardiac Stress Testing System; Programmable Diagnostic Computer

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.
Products
  • Hillrom Welch Allyn XScribe Cardiac Stress Testing System; Programmable Diagnostic Computer
Where it was sold

Worldwide distribution.

Nationwide

Barcodes
007320942590320073209425907000732094259100007320942591480073209425917900732094259230007320942592610073209425928500732094259292007320942593080073209425931500732094259322007320942593460073209425944500732094259506007320942595750073209425961200732094259650007320942596810073209425973500732094259766007320942598100073209425988900732094259919007320942599710073209426012000732094260137007320942601510073209426019900732094260229007320942602670073209426030400732094260342007320942603800073209426043400732094260519007320942605330073209426055700732094260601007320942606700073209426073100732094260816007320942608470073209426088500732094260892007320942609600073209426099100732094261004007320942610280073209426104200732094261196007320942942550073209429453800732094294545007320942954290073209429630300732094299908007320942999150073209430147200732094302974007320943169260073209432017600732094320183007320943327800073209433751800732094338805007320943394440081234502755600817655021997
Units affected
766 units