Baxter Healthcare Corporation recalls Hillrom Welch Allyn XScribe Cardiac Stress Testing System; Programmable Diagnostic Computer
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.
- Products
-
- Hillrom Welch Allyn XScribe Cardiac Stress Testing System; Programmable Diagnostic Computer
- Where it was sold
-
Worldwide distribution.
Nationwide
- Barcodes
007320942590320073209425907000732094259100007320942591480073209425917900732094259230007320942592610073209425928500732094259292007320942593080073209425931500732094259322007320942593460073209425944500732094259506007320942595750073209425961200732094259650007320942596810073209425973500732094259766007320942598100073209425988900732094259919007320942599710073209426012000732094260137007320942601510073209426019900732094260229007320942602670073209426030400732094260342007320942603800073209426043400732094260519007320942605330073209426055700732094260601007320942606700073209426073100732094260816007320942608470073209426088500732094260892007320942609600073209426099100732094261004007320942610280073209426104200732094261196007320942942550073209429453800732094294545007320942954290073209429630300732094299908007320942999150073209430147200732094302974007320943169260073209432017600732094320183007320943327800073209433751800732094338805007320943394440081234502755600817655021997- Units affected
- 766 units
- About this recall
- Hillrom Welch Allyn XScribe Cardiac Stress Testing System; Programmable Diagnostic Computer