Sentinel CH SpA recalls Alinity c Iron Reagent
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.
- Products
-
- Alinity c Iron Reagent
- Company
- Sentinel CH SpA
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
00380740136253- Units affected
- 4400 kits
- About this recall
- Alinity c Iron Reagent, Reference Number 08P3920