restor3d Inc. recalls Veritas Reverse Total Shoulder Replacement
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- (Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.
- Products
-
- Veritas Reverse Total Shoulder Replacement
- Company
- restor3d Inc.
- Where it was sold
-
US Nationwide distribution in the states of Texas, Massachusetts, Oklahoma, Florida, Colorado, North Carolina.
States: Colorado, Florida, Massachusetts, North Carolina, Oklahoma, Texas
- Barcodes
00840097513492008400975134610084009751343000840097513409- Units affected
- 240 units
- About this recall
- Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner