BioFire Diagnostics, LLC recalls FilmArray TORCH REF HTFA-ASY-0104. HTFA-ASY-0104 is the BIOFIRE TORCH Base which includes barcode scanner
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Due to an increased risk of degradation of the power entry switch resulting from arcing inside of the power switch that may result in carbon build-up on the switch contacts. The carbon build-up may lead to excess heating inside of the power switch, subsequently leading to deformation of the power switch case. The deformation can create an opportunity for a power switch electrical short. In addition to an electrical short, deformation may result in an open circuit causing the power switch to fail. This event would only occur after the product is in use.
- Products
-
- FilmArray TORCH REF HTFA-ASY-0104. HTFA-ASY-0104 is the BIOFIRE TORCH Base which includes barcode scanner
- Company
- BioFire Diagnostics, LLC
- Where it was sold
-
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, N, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, Virgin Island, VT, WA, WI, WV, and WY. The countries of Algeria, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Egypt, El Salvador, France, Germany, Guatemala, India, Indonesia, Italy, Japan, Korea, Kuwait, Malaysia, Mexico, New Zealand, Paraguay, Philippines, Poland, Qatar, Romania, Saudi Arabia, Singapore, Spain, Taiwan, Turkey, United Arab Emirates.
Nationwide
- Barcodes
00815381020253- Units affected
- 5,406 systems
- About this recall
- FilmArray TORCH REF HTFA-ASY-0104. HTFA-ASY-0104 is the BIOFIRE TORCH Base which includes barcode scanner, touch screen computer and USB ports. The BioFire FilmArray Torch (BioFire Torch) is an automated in vitro diagnostic (IVD) device intended for use with FDA-cleared or approved IVD BioFire FilmArray Panels. Additional Part/Reference Numbers as part of the system: - HTFA-ASY-0102: BIOFIRE TORCH Duplex. This is a plastic component that covers the BIOFIRE TORCH Modules. The BIOFIRE TORCH Duplex is only necessary if the system includes more than two modules. - HFTA-SUB-0103: BIOFIRE TORCH Module. The BIOFIRE reagent pouches are inserted into the modules.