Baxter Healthcare Corporation recalls Operating Table TruSystem 7000 U (dV)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.
- Products
-
- Operating Table TruSystem 7000 U (dV)
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
00887761967519- Units affected
- 62 units
- About this recall
- Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633