Stryker Corporation recalls Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
- Products
-
- Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil
- Company
- Stryker Corporation
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.
Nationwide
- Barcodes
07613327117202- Units affected
- 1,120,860 eaches
- About this recall
- Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-047-000