Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Stryker Corporation recalls Stryker Neptune SafeAir Smoke Evacuation Pencil Coated Rocker-Switch

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
Products
  • Stryker Neptune SafeAir Smoke Evacuation Pencil Coated Rocker-Switch
Where it was sold

Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.

Nationwide

Barcodes
07613327379471
Units affected
295,760 eaches
About this recall
Stryker Neptune SafeAir Smoke Evacuation Pencil Coated Rocker-Switch, Bulk Non-Sterile - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-047-002