Breas Medical, Inc. recalls with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Produced with unintended open slits on the side of the mouthpiece body, may result in insufficient ventilation in spontaneous breathing patients using MPV-VCV and MPV-PCV modes of ventilation
- Products
-
- with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093
- Company
- Breas Medical, Inc.
- Where it was sold
-
Nationwide Foreign: Canada
Nationwide
- Barcodes
05030267120505- Units affected
- 215 units
- About this recall
- with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093, 006653, an accessory to ventilator device, Vivo 45LS. intended for partial insertion into a patient's mouth (in between the lips) to facilitate access to the respiratory system.