Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

AVID Medical, Inc. recalls Halyard KNEE ARTHROSCOPY PACK - Medical convenience kits Model Number: MAMC204-16

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
Products
  • Halyard KNEE ARTHROSCOPY PACK - Medical convenience kits Model Number: MAMC204-16
Where it was sold

US Nationwide distribution.

Nationwide

Barcodes
10809160360767
Units affected
22 eaches
About this recall
Halyard KNEE ARTHROSCOPY PACK - Medical convenience kits Model Number: MAMC204-16