FDAMedical devicesRecalled Status: Ongoing

Beckman Coulter, Inc. recalls Beckman Coulter UniCel DxI 600

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.
Products
  • Beckman Coulter UniCel DxI 600
Where it was sold

Worldwide

Nationwide

Barcodes
1509959036922415099590341602
Units affected
80 units
About this recall