Medline Industries, LP recalls Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product n
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
- Products
-
- Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product n
- Company
- Medline Industries, LP
- Where it was sold
-
International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
Nationwide
- Barcodes
40193489400268401934894190004019532709255040195327478286401984593207064019348943078440198459320720408882779622014019348983205240193489603188408899426325354088994265543540888277761958408899428412274019348945552740195327279142401984593722934019348948598240193489250573401934893481024019348936533840193489429467401934894633864019845930188040195327298167401953273724234019532734103040195327694402401953275209924019532769550840195327214648- Units affected
- 96,406 total units
- About this recall
- Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA10591; ANES CIRC 120IN EXP-LF DYNJAA6555G; ANES CIRCUIT ADULT 120"-LF DYNJAA10817; ANESTHESIA 120" ADLT-LF DYNJAA10775; ANESTHESIA CIRC ADULT 120-LF DYNJAA10601; ANESTHESIA CIRCUIT - LF DYNJAA10261B; ANESTHESIA CIRCUIT 120EXP-LF DYNJAA10741A; ANESTHESIA CIRCUIT ADULT-LF DYNJAA0380A DYNJAA10641 DYNJAA10706B DYNJAA10851 DYNJAA10922B DYNJAA10971A DYNJAA11030A DYNJAA11087 DYNJAA11126; ANESTHESIA CIRCUIT-LF DYNJAA0309A DYNJAA0368A DYNJAA10368A DYNJAA10392 DYNJAA10430A DYNJAA10547A DYNJAA10639A DYNJAA10656A DYNJAA10685A DYNJAA10685B; CIRC 120" EXP 1 BV HMEF 3L GSL DYNJAA10920; EXPANDABLE CIRCUIT-LF DYNJAA11095; NM ANESTHESIA TURNOVER KIT DYNJAA0433; PROV HEALTH HEPA CIRC #2-LF DYNJAA0360; SPTMPX01IA CIRCUIT ADULT-LF DYNJAA10973A;