Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medline Industries, LP recalls Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; OPEN HEART SUPPLEMENT DYNJ909

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Products
  • Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; OPEN HEART SUPPLEMENT DYNJ909
Where it was sold

International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;

Nationwide

Barcodes
4019845905535640198459130619
Units affected
96,406 total units
About this recall
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; OPEN HEART SUPPLEMENT DYNJ909116D; VASCULAR DYNJ909124F