Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Elekta, Inc. recalls ELEKTA LIMITED Versa HD

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
Products
  • ELEKTA LIMITED Versa HD
Where it was sold

worldwide

Nationwide

Barcodes
0506019107152905060191071574
Units affected
451 units
About this recall
ELEKTA LIMITED Versa HD, REF XRT 2121, 2131; Medical Charged-Particle Radiation Therapy System Accelerator, Linear