Elekta, Inc. recalls ELEKTA AXESSE
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.
- Products
-
- ELEKTA AXESSE
- Company
- Elekta, Inc.
- Where it was sold
-
worldwide
Nationwide
- Units affected
- 47 units
- About this recall
- ELEKTA AXESSE, REF XRT 2181, XRT 27891; Medical Charged-Particle Radiation Therapy System Accelerator, Linear