Medline Industries, LP recalls Convenience Kits: DYNJ901048N ORTHO SPINE ST CLAIR DYNJ909123F UROLOGY MINOR DYNJ909132F SET UP DYNJ909308D HI

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Products
  • Convenience Kits: DYNJ901048N ORTHO SPINE ST CLAIR DYNJ909123F UROLOGY MINOR DYNJ909132F SET UP DYNJ909308D HI
Where it was sold

International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;

Nationwide

Barcodes
40198459055288401984590555544019845913057240198459130602
Units affected
96,406 total units
About this recall
Convenience Kits: DYNJ901048N ORTHO SPINE ST CLAIR DYNJ909123F UROLOGY MINOR DYNJ909132F SET UP DYNJ909308D HIP PINNING