Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Hobbs Medical, Inc. recalls Posi-Stop Injection Needle

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
Products
  • Posi-Stop Injection Needle
Where it was sold

Worldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN, MS, NH, NJ, NY, OH, TN, VT, WA, WI, WV and the country of Canada.

States: Connecticut, Georgia, Indiana, Michigan, Minnesota, Mississippi, New Hampshire, New Jersey, New York, Ohio, Tennessee, Vermont, Washington, Wisconsin, West Virginia

Units affected
198 units
About this recall
Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4722