Hobbs Medical, Inc. recalls Posi-Stop Injection Needle
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
- Products
-
- Posi-Stop Injection Needle
- Company
- Hobbs Medical, Inc.
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN, MS, NH, NJ, NY, OH, TN, VT, WA, WI, WV and the country of Canada.
States: Connecticut, Georgia, Indiana, Michigan, Minnesota, Mississippi, New Hampshire, New Jersey, New York, Ohio, Tennessee, Vermont, Washington, Wisconsin, West Virginia
- Units affected
- 198 units
- About this recall
- Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4722