NeuroLogica Corporation recalls GM85 Digital Mobile X-ray imaging System; Model Number: GM85;

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.
Products
  • GM85 Digital Mobile X-ray imaging System; Model Number: GM85;
Where it was sold

Domestic: US Nationwide Distribution. International: Australia, Belgium, Canada, Chile, China Mainland, Czech Republic, Denmark, Ecuador, Finland, France, Germany, GHANA, Hungary, India, Italy, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Saudi Arabia, Serbia, Singapore, Spain, Sweden, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab Emir., Vietnam,

Nationwide

Barcodes
08806088582412
Units affected
236 units
About this recall
GM85 Digital Mobile X-ray imaging System; Model Number: GM85;