Philips Respironics, Inc. recalls SimplyGo Mini
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Certain serial numbers of the SimplyGo Mini standard battery kit contain a defective chip which can lead the battery to either fail to charge or lose voltage output when in use in the device. If the fault occurs while the device is running on battery power, the SimplyGo Mini will shut down and stop providing oxygen to the user. This may happen at a temperature within the specified operating temperature range of the SimplyGo Mini.
- Products
-
- SimplyGo Mini
- Company
- Philips Respironics, Inc.
- Where it was sold
-
US Nationwide distribution in the states of GA, ID, KY, NJ, NY and the country of Canada.
States: Georgia, Idaho, Kentucky, New Jersey, New York
- Barcodes
00606959032385- Units affected
- 56 units (55 US, 1 OUS)
- About this recall
- SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113600