Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Becton Dickinson & Co. recalls BD Kiestra" Isolate Suspension Cuvette Array; Catalog No.: 246100;

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The 2D barcodes on these cuvettes contain formatting errors that prevent the analyzing system from correctly reading the expiration date. As a result, the instrument enters an error state and the cuvettes cannot be used.
Products
  • BD Kiestra" Isolate Suspension Cuvette Array; Catalog No.: 246100;
Where it was sold

Worldwide - US Nationwide distribution in the states of CO, CT, KY, NJ, TN, TX, UT and the countries of Belgium, Singapore, Taiwan, Germany, Switzerland.

States: Colorado, Connecticut, Kentucky, New Jersey, Tennessee, Texas, Utah

Barcodes
30382902461006
Units affected
29,760 units
About this recall