Becton Dickinson & Co. recalls BD Kiestra" Isolate Suspension Cuvette Array; Catalog No.: 246100;
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The 2D barcodes on these cuvettes contain formatting errors that prevent the analyzing system from correctly reading the expiration date. As a result, the instrument enters an error state and the cuvettes cannot be used.
- Products
-
- BD Kiestra" Isolate Suspension Cuvette Array; Catalog No.: 246100;
- Company
- Becton Dickinson & Co.
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of CO, CT, KY, NJ, TN, TX, UT and the countries of Belgium, Singapore, Taiwan, Germany, Switzerland.
States: Colorado, Connecticut, Kentucky, New Jersey, Tennessee, Texas, Utah
- Barcodes
30382902461006- Units affected
- 29,760 units
- About this recall
- BD Kiestra" Isolate Suspension Cuvette Array; Catalog No.: 246100;