Ortho-Clinical Diagnostics, INc. recalls VITROS Chemistry Products ECO2 Slides-For in vitro diagnostic use only. VITROS Chemistry Products ECO2 Slides
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Ortho Clinical Diagnostics (QuidelOrtho) confirmed during routine testing that some lots from Coating 3598 of VITROS ECO2 slides included carts which may contain consecutive slides that may not function as intended. These affected slides (potentially intermittent, consecutive slides within a cart or across consecutive carts) have an abnormality which may cause positively or negatively biased total carbon dioxide (CO2) results leading to either overdiagnosis of acidosis or missing alkalosis, both which may prompt unnecessary therapies such as IV fluids and/or bicarbonate or overdiagnosis of alkalosis which may also result in unnecessary use of IV fluids and electrolyte replacement or failure to detect acidosis and necessary treatment.
- Products
-
- VITROS Chemistry Products ECO2 Slides-For in vitro diagnostic use only. VITROS Chemistry Products ECO2 Slides
- Where it was sold
-
Nationwide Foreign: Bermuda Canada
Nationwide
- Barcodes
10758750004645- Units affected
- 622 units 300 Slides (5 cartridge containing 60 slides)
- About this recall
- VITROS Chemistry Products ECO2 Slides-For in vitro diagnostic use only. VITROS Chemistry Products ECO2 Slides quantitatively measure total carbon dioxide (CO2) concentration in serum and plasma using VITROS 250/350/5,1 FS/4600 XT 3400 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance. Product Code: 8262396