Medline Industries, LP recalls MEDLINE convenience kits labeled as: 1) MAJOR VAGINAL HARPER PACK-LF
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
- Products
-
- MEDLINE convenience kits labeled as: 1) MAJOR VAGINAL HARPER PACK-LF
- Company
- Medline Industries, LP
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Barcodes
1019532740456710195327228002- Units affected
- 424 kits
- About this recall
- MEDLINE convenience kits labeled as: 1) MAJOR VAGINAL HARPER PACK-LF, REF DYNJ0161768G; 2) ACH MAJOR VAGINAL PACK, REF DYNJ67757B.