Smiths Medical ASD, Inc. recalls smiths medical portex Venn Reusable Tracheal Tube Introducer
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.
- Products
-
- smiths medical portex Venn Reusable Tracheal Tube Introducer
- Company
- Smiths Medical ASD, Inc.
- Where it was sold
-
Worldwide distribution.
Nationwide
- Barcodes
35021239002579- Units affected
- 142327 units
- About this recall
- smiths medical portex Venn Reusable Tracheal Tube Introducer, 15Ch x 60cm, REF 14-504-17 and REF 14-504-17JP