Smiths Medical ASD, Inc. recalls smiths medical PORTEX Tracheal Tube Exchange Guide

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.
Products
  • smiths medical PORTEX Tracheal Tube Exchange Guide
Where it was sold

Worldwide distribution.

Nationwide

Barcodes
35021239002586
Units affected
10,223 units
About this recall
smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP