Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Ortho-Clinical Diagnostics, Inc. recalls VITROS 5600 Integrated System Product Code 6802413 and Refurbished Product Code 6802915 Running Software Versi

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.
Products
  • VITROS 5600 Integrated System Product Code 6802413 and Refurbished Product Code 6802915 Running Software Versi
Where it was sold

US Nationwide. Global Distribution.

Nationwide

Barcodes
1075875000274010758750007110
Units affected
176 units
About this recall
VITROS 5600 Integrated System Product Code 6802413 and Refurbished Product Code 6802915 Running Software Version 3.8.0