Baxter Healthcare Corporation recalls Baxter MiniCap Extended Life PD Transfer Set (short) with Titanium Spike

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
There has been an increase in the number of complaints related to the separation of the female connector from the main body of the transfer set.
Products
  • Baxter MiniCap Extended Life PD Transfer Set (short) with Titanium Spike
Where it was sold

Worldwide distribution.

Nationwide

Barcodes
0085412150703
Units affected
13376 units
About this recall
Baxter MiniCap Extended Life PD Transfer Set (short) with Titanium Spike, Part Number T5C4326K; use in Peritoneal Dialysis