Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Karl Storz Endoscopy recalls Injection Needle

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The efficacy of the manual cleaning process cannot be assured for the affected injection needles intended for general surgery and visceral surgery, which if used, may expose the patient to a higher risk of infection.
Products
  • Injection Needle
Where it was sold

US: MI, NH, SC, PA, FL, NY, NC, MS, MO, CA, NJ, IL, NV, OK, CO, OH, WA, WI, VA, IN, IA, AR, TX, WV, MD, MN, GA, MA, NM, VT, AK, OR, ID, LA, KS, DC, KY, HI OUS: Mexica, Canada, Germany

States: Alaska, Arkansas, California, Colorado, District of Columbia, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Mississippi, North Carolina, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, Texas, Virginia, Vermont, Washington, Wisconsin, West Virginia

Units affected
990
About this recall
Injection Needle, 3 Fr., REF: 27030N; Injection Cannula, 6 Fr., REF: 27182A