Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Pivotal Health Solutions, Inc. recalls Quantum Intersegmental Table

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.
Products
  • Quantum Intersegmental Table
Where it was sold

Distribution to US nationwide, Canada and Japan.

Nationwide

Units affected
464 units
About this recall
Quantum Intersegmental Table, Model/Catalog Number: AMQ400H; heated massage table