Zeus Scientific, Inc. recalls ZEUS ELISA Parvovirus B19 IgM Test System
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The organization completed a product improvement process to the conjugate found within the certain ELISA kits. This modification was not reviewed and approved by FDA prior to its implementation.
- Products
-
- ZEUS ELISA Parvovirus B19 IgM Test System
- Company
- Zeus Scientific, Inc.
- Where it was sold
-
US States: CA, MA, MI, NJ, NY, TN, UT.
States: California, Massachusetts, Michigan, New Jersey, New York, Tennessee, Utah
- Barcodes
00845533001724- Units affected
- 293 kits
- About this recall
- ZEUS ELISA Parvovirus B19 IgM Test System, Product Number SM9Z7701G