Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

B Braun Medical Inc recalls Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Products
  • Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's
Where it was sold

Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.

Nationwide

Barcodes
0404696499127104046964991264
Units affected
36,552 units
About this recall
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML, 134 IN.; Catalog Number: 490477.