Atrium Medical Corporation recalls Atrium Oasis Drain Single W / AC
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- A sterilization nonconformance prematurely aged the product by 48.3 days, resulting in the expiration date on the label not being accurate. This nonconforming product was inadvertently distributed to customers. Increased risk to patients and healthcare providers using the product after the shortened expiration date of November 17, 2025.
- Products
-
- Atrium Oasis Drain Single W / AC
- Company
- Atrium Medical Corporation
- Where it was sold
-
Domestic distribution nationwide. No foreign distribution.
Nationwide
- Barcodes
00650862110012- Units affected
- 835 cases total (6 units/case)
- About this recall
- Atrium Oasis Drain Single W / AC, Part No. 3600-100