Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

B Braun Medical Inc recalls IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BB

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Products
  • IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BB
Where it was sold

Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.

Nationwide

Barcodes
04046955203284040469552032770404695527034704046955270330040469553528760404695535286904046955664696040469556646890404695583636904046955836352
Units affected
190,625 units
About this recall
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15 DROP W/ CARESITE EXT SET; Catalog Number: 490528. 2. 111 IN. IV ADMIN SET W/ 12 IN. Y-EXT SET; Catalog Number: 490550. 3. IV ADMIN SET, 125 IN. w/ EXT SET 9 IN.; Catalog Number: 490605. 4. IV SET 15 DR 3CARESITE 2 ULTRAPORT; Catalog Number: 490616. 5. IV ADMIN SET, 2 CARSITE LADS; Catalog Number: 490654.