Philips North America recalls BrightView
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients may include fracture, body part loss of function/debilitation, muscle or ligament sprain or strain, laceration, crush injury, abrasion, or contusion.
- Products
-
- BrightView
- Company
- Philips North America
- Where it was sold
-
Domestic distribution nationwide. Foreign distribution worldwide.
Nationwide
- Units affected
- 359 US; 328 OUS
- About this recall
- BrightView, Gamma Camera System, Product Code 882480.