Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Diasorin Inc. recalls DiaSorin LIAISON Q.S.E.T. Device Plus

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Customer complaints were received alleging leaking devices and insufficient buffer volume. Diasorin's internal investigation determined that 0.14% of the LIAISON Q.S.E.T. Device Plus in the lots identified have loose clear caps, which may allow buffer to leak from the tube. A stool sample prepared with less than the correct amount of buffer may result an incorrectly high patient result.
Products
  • DiaSorin LIAISON Q.S.E.T. Device Plus
Where it was sold

Worldwide

Nationwide

Units affected
21905 units
About this recall
DiaSorin LIAISON Q.S.E.T. Device Plus, REF 319060; intended for use in the preparation of human stool specimens