Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Philips North America recalls BrightView XCT

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients may include fracture, body part loss of function/debilitation, muscle or ligament sprain or strain, laceration, crush injury, abrasion, or contusion.
Products
  • BrightView XCT
Where it was sold

Domestic distribution nationwide. Foreign distribution worldwide.

Nationwide

Units affected
135 US; 294 OUS
About this recall
BrightView XCT, Gamma Camera, Product Code 882482