Philips North America Llc recalls Philips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The following four issues are included: 1. Certain Channel Settings may result in an Access Point preventing a connected client device from roaming to another Access Point potentially causing the client device to appear to drop out from the network. 2. if the Access Point experiences a disconnection from the Access Point Controller exceeding 4 minutes, the Access Point may not transition through the desired state to all for clean reconnection of the client radio link. 3. Potential to experience higher than expected data dropouts, device disassociations/reassociations for their wireless clients, and frequent "Loss of AP" alerts. 4. Potential for the Smart-hopping Access Point is affected by a software issue where the Access Point will resent after 82.85 days when the frame counter reaches its limit. This causes the latest synchronization check to fail and potential for loss of patient data.
Products
  • Philips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216
Where it was sold

Nationwide distribution

Nationwide

Barcodes
00884838099876
Units affected
4,811 units
About this recall
Philips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216