Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Icotec Ag recalls VADER pedicle system

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.
Products
  • VADER pedicle system
Company
Icotec Ag
Where it was sold

US: AZ. MI, OH, CA, CO, TX, NY, NC, DE, PA

States: Arizona, California, Colorado, Delaware, Michigan, North Carolina, New York, Ohio, Pennsylvania, Texas

About this recall
VADER pedicle system, torque wrench, Catalog Number 42-703