Icotec Ag recalls VADER pedicle system
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.
- Products
-
- VADER pedicle system
- Company
- Icotec Ag
- Where it was sold
-
US: AZ. MI, OH, CA, CO, TX, NY, NC, DE, PA
States: Arizona, California, Colorado, Delaware, Michigan, North Carolina, New York, Ohio, Pennsylvania, Texas
- About this recall
- VADER pedicle system, torque wrench, Catalog Number 42-703